Specific Safety Instructions of High Speed Centrifuge 04
Standard Operation
15. At end of run, do not open chamber until the rotor comes to rest.
16. After rotor has stopped, press VACUUM to vent chamber.
17. When working with infectious materials wait 10 minutes after the rotor has stopped
before opening the centrifuge lid. Be aware of decontamination and clean-up procedures that apply to materials being centrifuged.
18. If a breakage has occurred, wait at least 10 minutes after the rotor has stopped for
aerosols and droplets to settle before opening lid.
19. Open chamber door and remove the rotor. Ensure rotor chamber is clean and dry.
Keep door closed between runs.
20. Centrifugation causes a slight vacuum to build up in the tube cavity, occasionally
resulting in the suction effect when removing the tubes from the rotor. Allow at least 5
minutes after the rotor comes to rest before removing the tubes from the rotor.