Kaida Large Capacity and High Speed Centrifuge (bench top) Standard Operation Part 1

 

Standard Operation

 

1. Turn power switch on.

 

2. Use only Kaida rotors and accessories designed for use in this instrument.

 

3. Inspect Kaida rotor for signs of corrosion or cracking before using. If found, do not use the rotor, and inform the person responsible for the centrifuge of the problem.

 

4. Inspect tubes and bottles for cracks or flaws prior to use. Do not use if damaged.

 

5. Do not interchange tube-cap components. They are designed as a unit for a particular tube being centrifuged in a particular rotor.

 

6. Minimize the differences in weight of the filled sample tubes. Refer to instruction manual for details.

 

7. The maximum recommended weight of a filled bucket is stated on each rotor.

 

8. Excessively low liquid level may limit the rotor speed or crush the tube. Refer to instruction manual for details.

 

9. To avoid corrosion do not expose aluminum part of rotor to strong acids, bases or alkaline laboratory detergents, liquid chlorine bleach or salts of heavy metals. If exposure does occur, clean the rotor immediately.

 

10. Rotation speed must be reduced below the maximum for liquids with a density of 1.2g/ccm or higher.

 

11. Before running a swinging rotor, make sure that each is mounted correctly and that they can swing freely. Ensure grooves are clean before inserting buckets. Unclean grooves and pivots prevent buckets from swinging freely.