13. Install rotor. Close and lock lid.
14. Enter the required run conditions.
15. Press START key. Run operates for specified operation time or press STOP key to
stop.
16. Do not use the lid closure override to open the instrument before the rotor has
stopped. Wait until the rotor has stopped before opening the lid to remove samples.
17. When working with infectious materials wait 10 minutes after the rotor has stopped
before opening the centrifuge lid. Be aware of decontamination and clean-up procedures
which apply to materials being centrifuged.
18. If a breakage has occurred, wait at least 10 minutes after the rotor has stopped for
aerosols and droplets to settle before opening lid.
19. Centrifugation may cause a slight vacuum to build up in the tube cavity, occasionally
resulting in the suction effect when removing the tubes from the rotor. When this occurs,
allow at least 5 minutes after the rotor comes to rest before removing the tubes from the